A1: the full patient identifier is (b)(6).A2, a3, a4 and a5: the customer did not supply patient demographics such as age, date of birth, sex, weight, ethnicity or race.H3 and h6: the access total t3 reagent was not returned for evaluation.No hardware errors, flags were reported in conjunction with this event.No issues with samples integrity were reported by the customer.Service was dispatched.Field service engineer (fse) replaced reagent and calibrator pack.The fse also tested some samples using heterophile blocking tubes but the results were unchanged.In conclusion, the cause of this event cannot be determined with the available information.There is no evidence to reasonably suggest that a malfunction occurred in conjunction with this event.Note 1: a second medwatch report will address the questioned free t4 results for this patient.Note 2: a second patient with questioned free t4 and total t3 results was reported on (b)(6) 2024.Two other medwatch reports will address the questioned total t3 and free t4 results for this second patient.
|