| Catalog Number |
365716 |
| Medical Device Problem Code |
Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
04/24/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that when attempting to remove foley nursing staff was unable to deflate balloon.Urology was called to the bedside and was able to remove but this caused significant trauma to the urethra and bleeding from the meatus.It was unknown what medical intervention was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that when attempting to remove foley nursing staff was unable to deflate balloon.Urology was called to the bedside and was able to remove but this caused significant trauma to the urethra and bleeding from the meatus.Per follow up via email on 20may2024, this was the only lot that they had this particular issue with.Outside of urethral trauma at time of removal there had been no reported patient impact associated with this.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive because no sample was returned.It was unknown whether the device had met relevant specifications.The product was used for urological care.It was unknown whether the product had caused the reported failure.The potential root cause for this failure could be user related (example: over aspirated, incorrect syringe/collapse lumen/sac close eye/valve damage).A potential root cause for this failure mode could be, thin rubberize wall that causing collapse/pinch inflation lumen under the balloon or along the shaft.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.The instructions for use were found adequate and state the following: "do not use ointments or lubricants having a petrolatum base.They will damage latex and may burst balloon.Caution: this product contains natural rubber latex which may cause allergic reactions.Sterile unless package is opened or damaged.Do not use if package is opened or damaged.Warning: on catheter, do not use ointments or lubricants having a petrolatum base.They will damage latex and may cause balloon burst.Balloon burst can lead to premature deflation."visually inspect the product for any imperfections or surface deterioration prior to use.Recommended inflation capacities 3cc balloon: use 5cc sterile water 5cc balloon: use 10cc sterile water 15cc balloon: use 20cc sterile water 20cc balloon: use 25cc sterile water 30cc balloon: use 35cc sterile water 40cc balloon: use 45cc sterile water 75cc balloon: use 50cc sterile water do not exceeded recommended capacities." to deflate catheter balloon: gently insert a syringe in the catheter valve.Never use more force than is required to make the syringe "stick" in the valve.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact and adequately trained professional for assistance, as directed by hospital protocol." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|