| Model Number |
EGIATRS45AMT |
| Medical Device Problem Codes |
Material Rupture (1546); Failure to Form Staple (2579)
|
| Health Effect - Clinical Code |
Tissue Breakdown (2681)
|
| Date of Event |
04/10/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Concomitant product: sigtrsb45amt 45mm med thk reinforced rel (lot#: n2m0301y) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, in a laparoscopic bariatric procedure, when firing the stapler in the stomach, the reinforcement sheet of two reloads tore during stapling.There was poor staple formation.It was necessary to use additional reloads to resolve the poor staple formation.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6 correction: e1 (fax number) h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but a photo was available for evaluation.Visual inspection noted that there were four images of staple lines.The staples appeared to have properly formed.Two pieces of torn reinforcement material could be seen outside of the tissue cavity.It was reported that there was poor staple formation.The reported issue could not be confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|