(b)(4).The optiflow junior nasal cannula is designed specifically for the delicate anatomical features and flow requirements of neonatal and pediatric patients.It features adhesive pads (wagglepads) to maintain cannula stability on the patient's cheeks, soft-touch nasal prongs, and breathable kink-proof, and crush-resistant flexible tubing.Method: seven complaint ojr416b junior 2 nasal cannula were received at fisher & paykel healthcare (f&p) new zealand.It was further reported that 3 cannulas were not in the original packaging while 4 cannulas were in their original packaging and sealed.The devices were visually inspected and chemically analyzed.Results: visual inspection of the complaint devices revealed that 3 cannulas had holes in the tubing while no fault found on four cannulas from sealed packaging.Chemical analysis confirmed that the tubing was wiped with or exposed to a chemical resulting in the material dissolving and forming holes.Conclusion: we are unable to determine the cause of the reported event.However, the results from the chemical analysis supported the conclusion that the hole formation of the tube material was likely due to chemical exposure.All optiflow junior cannulas are 100% leak and occlusion tested after final assembly and any cannula that fails is discarded.Samples are also taken and pull tested to check the tube tensile strength and the glue joint strength at the cannula/ tube joint, as well as the swivel grip joint.The subject cannula would have met the required specification at the time of production.The user instructions illustrate in pictorial format the correct set-up and proper use of the optiflow junior 2 nasal cannula.They also state the following: appropriate patient monitoring (e.G.Oxygen saturation) must be used at all times.Failure to monitor the patient (e.G.In the event of an interruption to gas flow) may result in serious harm or death.Do not soak, sterilize, or reuse this product.Avoid contact with chemicals, cleaning agents, or hand sanitizers.Secretions on the cannula and the prongs can be removed by gently wiping with a damp cloth.Do not wrap, insulate, stretch or crush the tubing as this may impair the performance of this product or compromise safety including potentially causing patient harm).Ensure all connections are secure during use.Check the cannula is undamaged and that the flow path is maintained.Under excessive load, the cannula may disconnect to prevent forces being transferred to the patient.Ensure the patient does not lie on the tubing a this may apply pressure to the patient's ears or face.Failure to apply and use this product within the directions, transport, storage and operating conditions specified in the labelling and user instructions may impair performance of this product or compromise safety (including potentially causing serious patient harm.).
|