This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.A review of device history record and historical complaints analysis was conducted and no deviations that could have caused or contributed to the reported issue was identified.Based on the results of the investigation, a definitive root cause could not be determined.However, it is likely that "coating on the insertion tube of the insertion section was peeled off" occurred due to stress of repeated use, external factors, or handling of the device.The event can be detected/prevented by following the instructions for use which state: 3.2 inspection of the endoscope.5.Inspect the external surface of the entire insertion tube including the bending section and the distal end for dents, bulges, swelling, peeling, scratching, holes, sagging, transformation, bends, adhesion of foreign body, dropout of parts, any protruding objects or other irregularities.Olympus will continue to monitor field performance for this device.
|