| Model Number |
Q919 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/07/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been returned.Production records were reviewed.There were no nonconformities noted with the lot during production.All finished goods testing requirements were met prior to release.Three retained samples from the same lot were used to perform a needle attachment test.The retained samples used for the needle attachment test met requirements.There was no evidence that the device failed to meet specifications, and the report could not be substantiated.A cause for the event cannot be established.This report and use of categorical definitions required by fda 3500a does not constitute an admission by riverpoint medical or its employees that riverpoint medical or its employees have caused or contributed to the event described in this report.Riverpoint medical has filed this information to comply with the medical device reporting regulation 21 cfr 803.If additional information is provided to riverpoint medical regarding this event, a supplementary 3500a form will be submitted as required by the fda.
|
| |
|
Event or Problem Description
|
|
According to the reporter: "salvin 5-0 pgcl violet monoderm suture-fs-2 (c6) needle-18in.We received a complaint where the customer experienced, the needle fell off.The suture is from lot 24010406.".
|
| |
|