SMITH & NEPHEW, INC. 3.0MM CANNULATED SCREW 40MM; PLATE, FIXATION, BONE Back to Search Results
Catalog Number 71103240S
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/19/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during the osteosynthesis surgery, the head of one (1) 3.0mm cannulated screw 36mm and one (1) 3.0mm cannulated screw 40mm were broken.The procedure was resumed, after a non-significant delay, with a s+n back-up device.No injury was reported as a consequence of this issue.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
The device was not returned for evaluation; therefore, a device analysis could not be performed.The clinical/medical investigation concluded that, per complaint details, the head of the screw broke during screw placement of the osteosynthesis procedure.Reportedly, the procedure was resumed after a non-significant delay using a smith and nephew back-up device with no patient injury.The requested clinical information has not been provided as of the date of this medical investigation; therefore, definitive contributing clinical factors cannot be concluded.With the limited information provided, the patient impact beyond the reported non-significant delay and use of a back-up device cannot be determined.Device batch number was not provided, thus, an evaluation of the manufacturing records could not be performed.A review of complaint history for the previous 12 months did not reveal similar events for the listed device.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.According with material specification, shall control the quality and manufacture of special quality stainless steel cannulated bar.The material can be used for surgical implants and can considered to be surgical grade stainless steel.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.Factors that could contribute to the reported event include size selected, surgical technique, and/or excessive forces.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Additional Manufacturer Narrative
The associated device was returned and evaluated.The visual inspection revealed that the head of the screw broke off.This piece was not returned.An assessment of material characteristics concluded that the head was broken off during surgery.Aside from the head, no other fractures or defects were seen on the screw.It is not clear what initiated the screw breakage during surgery.The clinical/medical investigation concluded that the requested clinical information has not been provided as of the date of this medical investigation; therefore, definitive contributing clinical factors cannot be concluded.With the limited information provided, the patient impact beyond the reported non-significant delay and use of a back-up device cannot be determined.Device batch number was not provided, thus, an evaluation of the manufacturing records could not be performed.A review of complaint history for the previous 12 months did not reveal similar events for the listed device.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.According with material specification, the quality and manufacture of special quality stainless steel shall be controlled.This material can be used for surgical implants and can be considered surgical grade.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.Factors that could contribute to the reported event include size selected, surgical technique, and/or excessive forces.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3.0MM CANNULATED SCREW 40MM
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key19321363
Report Number1020279-2024-00939
Device Sequence Number7659703
Product Code HRS
UDI-Device Identifier03596010623997
UDI-Public03596010623997
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K210837
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number71103240S
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/19/2024
Supplement Date Received by Manufacturer06/19/2024
11/08/2024
Initial Report FDA Received Date05/15/2024
Supplement Report FDA Received Date06/20/2024
11/11/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age38 YR
Patient SexMale
-
-