| Model Number |
6944-58 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Over-Sensing (1438)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/20/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient presented to the hospital due to a lead integrity alert (lia) triggered on the right ventricular (rv) lead with over one-hundred short v-v intervals and non-sustained ventricular tachycardia episodes and noise was noted.It was further reported that there was electromagnetic interference (emi)/noise on the implantable cardioverter defibrillator (icd) and a loose set screw issue was suspected.The rv lead and icd remain in use.No patient complications have been reported as a result of this event.
|
| |
|