|
This device is classified as import for export, therefore 510k is not applicable.Model eg36-j10ur-us is available in the usa with a 510k number k200090.The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the air tube clogged.Based on the result, we concluded that it was caused due to the inadequate/insufficient reprocessing at the facility on the air tube.In addition, our technician confirmed that the air nozzle clogged, the water nozzle clogged, the water tube clogged, the lg prong corroded, the lg prong top corroded, and the remote control buttons cut; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0630(air/water & jet water channels)"" and/or the risk analysis results, it was evaluated to submit mdr.
|