• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TORNIER INC TORNIER PERFORM REVERSED PERIP SCREW 5.0MMX38MM NON STERILE; SHOULDER PROSTHESIS, REVERSE CONFIGURATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TORNIER INC TORNIER PERFORM REVERSED PERIP SCREW 5.0MMX38MM NON STERILE; SHOULDER PROSTHESIS, REVERSE CONFIGURATION Back to Search Results
Catalog Number DWJ338
Medical Device Problem Codes Break (1069); Fracture (1260); Migration (4003)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 03/06/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device will not be returned.If additional information becomes available, it will be provided on a supplemental report.
 
Event or Problem Description
The primary reverse procedure employed tornier implants, including the shoulder id baseplate and perform humeral.The observed baseplate failure is attributed to likely over-lateralization, stemming from the thickness of the shoulder id baseplate and a lateralized glenosphere.The screw also had a breakage.
 
Additional Manufacturer Narrative
The reported event of implant migration and screw breakage can be confirmed could be confirmed since x-rays were provided.A device inspection was not possible since the affected device was not returned.Since x-rays were provided, the opinion of the medical expert was requested and stated as following: whether the implant was correctly planned and correctly implanted cannot be assessed from the x-ray showing the implant migration alone.The event of implant migration and screw breakage can be confirmed.Screw breakage is secondary to implant migration and not a root cause.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history was not possible because the lot number was not communicated.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.Indications of material, manufacturing, or design related problems were unable to be identified as the lot number was not communicated.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Event or Problem Description
The primary reverse procedure employed tornier implants, including the shoulder id baseplate and perform humeral.The observed baseplate failure is attributed to likely over-lateralization, stemming from the thickness of the shoulder id baseplate and a lateralized glenosphere.The screw also had a breakage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TORNIER PERFORM REVERSED PERIP SCREW 5.0MMX38MM NON STERILE
Common Device Name
SHOULDER PROSTHESIS, REVERSE CONFIGURATION
Manufacturer (Section D)
TORNIER INC
10801 nesbitt avenue s
bloomington MN 55437
Manufacturer (Section G)
WRIGHT MEDICAL CORK (TORNIER ORTHOPEDICS IRELAND LTD)
harnetts cross
macroom, co. cork NA
EI   NA
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key19328764
Report Number0001649390-2024-00234
Device Sequence Number20560698
Product Code PHX
UDI-Device Identifier00846832089307
UDI-Public00846832089307
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K161742
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2024
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue NumberDWJ338
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/18/2024
Supplement Date Received by Manufacturer07/16/2024
Initial Report FDA Received Date05/16/2024
Supplement Report FDA Received Date08/12/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
Patient SexMale
-
-