| Model Number |
GALILEO® CAPTURING ROD |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/21/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.The complaint device was not received for evaluation.Based off the information provided, the most likely cause for the reported failure can be attributed to misuse/mishandling due to damage to the device.
|
| |
|
Event or Problem Description
|
|
On 01/21/2022 it was reported by a sales representative via sems that the 5033-100, galileo lag screw capturing rod is stuck in the 5046-100 capturing rod nut and would come out causing both the rod nut and the rod to be stripped.No patient affect reported.There was no additional information provided.
|
| |
|