| Model Number |
SUTURE LASSO, PIGTAIL, LEFT CURVED |
| Medical Device Problem Code |
Failure to Advance (2524)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
04/22/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On 26 april 2024, it was reported by warehouse via email that for an ar-8690sll suture lasso, pigtail, left curved the surgeon was unable to pass the wire loop through the pigtail lasso.This occurred during a plantar plate repair in southern cross hospital, christchurch.The case was completed successfully via the use of another suture lasso.There was case involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error of the device due to excessive force being used when attempting to advance the wire.
|
| |
|