EXACTECH, INC. LOGIC TIBIA IMPLANT PSC INSERT, SZ 3.5, 13MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
02-012-44-3513 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
D10: concomitant devices: (b)(6) 200-02-32 - three peg patella 32mm.(b)(6) 02-010-01-0335 - logic femoral ps cem right sz 3.5.(b)(6) 02-012-45-3535 - lgc tibial fit tray cem sz 3.5f / 3.5t.The device was not returned for evaluation and no photographs or x-rays were provided for review; therefore the reported event cannot be confirmed through analysis.No operative notes were provided; therefore, a review of device usage and technique could not be performed.No information concerning patient conditions, co-morbidities, or other health or anatomical concerns was provided and it is therefore unknown if patient-related factors may have caused or contributed to the reported event.The cause of the reported event cannot be determined based on the information made available.Should additional, relevant information become available, a supplemental report will be filed accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported via legal documentation that approximately 116 months after a right total knee replacement procedure, the patient had a planned revision procedure due to prosthesis wear.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: corrected the following: health effect - type of investigation.The reason for the reported event cannot be confirmed from the information provided but may be due to prosthesis wear and/or inclusion of the polyethylene in the packaging recall.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images, radiographs, and relevant clinical information were not provided.
|
| |
|