• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COLOPLAST A/S X-FLOW PROSTATECTOMY CATHETER; INDWELLING URETHRAL DRAINAGE BALLOON CATHETER, NON-ANTIMICROBIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COLOPLAST A/S X-FLOW PROSTATECTOMY CATHETER; INDWELLING URETHRAL DRAINAGE BALLOON CATHETER, NON-ANTIMICROBIAL Back to Search Results
Catalog Number AB6118
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code Discomfort (2330)
Date of Event 04/04/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The lot number was reviewed for complaint trend, nonconforming report and capa.Nc (b)(4): "pb ballons (bulles + agglutinés)" opened in (b)(6) 2023.For this nc, the used of clumped balloons should be responsible of some weakness on the balloon and being responsible of bursting.Monitoring capa-000152: "balloon issues on folysil and silicone prostatic catheters".The trending for balloon bursting is specifically monitored.
 
Event or Problem Description
According to the available information when checking the correct positioning of the balloon, the balloon was inflated to 50cc when the balloon burst.Patient experienced pain and non-irrigated hematuria.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X-FLOW PROSTATECTOMY CATHETER
Common Device Name
INDWELLING URETHRAL DRAINAGE BALLOON CATHETER, NON-ANTIMICROBIAL
Manufacturer (Section D)
COLOPLAST A/S
1 holtedam humlebaek, dk 3050
humlebaek 3050
DA  3050
Manufacturer Contact
usbes brian schmidt
1601 west river road n
minneapolis, MN 55411
MDR Report Key19334928
Report Number9610711-2024-00116
Device Sequence Number3009691
Product Code EZL
UDI-Device Identifier03600040143629
UDI-Public03600040143629
Combination Product (Y/N)N
Initial Reporter CountryFR
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 05/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberAB6118
Device Lot Number9453599_AB61181002
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date05/16/2024
Date Device Manufactured11/08/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-