BOSTON SCIENTIFIC CORPORATION AMPLATZ SUPER STIFF; WIRE, GUIDE, CATHETER Back to Search Results
Model Number 39575
Medical Device Problem Codes Stretched (1601); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/04/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluated by mfr.: the guidewire was returned for analysis.Visual inspection was performed which revealed the guidewire was bent and stretched at the distal tip section.Additionally, the core wire was exposed due to core wire separation from the bald solder.Based on analysis of the returned device, the reported guidewire stretching was confirmed.
 
Event or Problem Description
Reportable based on device analysis completed on (b)(6) 2024.It was reported that guidewire stretched occurred.An amplatz super stiff guidewire was selected for use.Upon unpacking, it was found that the guidewire was stretched and could not be used normally.The procedure was completed with another of the same device.There were no patient complications reported.However, returned device analysis revealed core wire separation from bald solder.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZ SUPER STIFF
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key19336866
Report Number2124215-2024-29712
Device Sequence Number6519155
Product Code DQX
UDI-Device Identifier08714729115830
UDI-Public08714729115830
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K843016
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/14/2026
Device Model Number39575
Device Catalogue Number39575
Device Lot Number0031624517
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/11/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/24/2024
Initial Report FDA Received Date05/16/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/15/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age69 YR
Patient SexFemale
Patient Weight68 KG
-
-