| Model Number |
M00545150 |
| Medical Device Problem Codes |
Break (1069); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/24/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Block d4, h4: the complainant was unable to provide the suspect device lot number; therefore, the lot expiration and device manufacture dates are unknown.Block h6: imdrf device code a0401 captures the reportable event of cutting wire broken.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that an autotome rx 49 was used in the duodenal papilla during an endoscopic papillotomy procedure performed on (b)(6) 2024.During the procedure, the cutting wire of the rx 49 broke off after stepping on the foot pedal several times during an attempt to perform a papillotomy.It was reported that no part of the cutting wire detached and fell into the patient.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Block d4, h4: the complainant was unable to provide the suspect device lot number; therefore, the lot expiration and device manufacture dates are unknown.Block h6: imdrf device code a0401 captures the reportable event of cutting wire broken.Block h11: the returned autotome rx 49 was analyzed, and a visual evaluation noted that the cutting wire was blackened, kinked and broken from the proximal pierced hole.No other problems with the device were noted.The reported event of cutting wire break was confirmed.Upon analysis, it was found that the cutting wire was kinked and broken from the proximal pierce hole.Based on the condition of the device, the problem found could have been generated if there was contact between the device and the scope during energization or if the generator exceeded the maximum of voltage during the procedure.Also, energizing the device prior to performing sphincterotomy can compromise the cutting wire's integrity and cause a premature cutting wire fatigue.Once the cutting wire breaks, any attempt to remove the device from the scope can lead to the broken section hitting the working channel of the scope.This can kink the cutting wire.It is most likely that procedural or anatomical factors encountered during the use of the device could have affected the device performance and its integrity.Based on all gathered information, the most probable root cause of this complaint is adverse event related to procedure.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that an autotome rx 49 was used in the duodenal papilla during an endoscopic papillotomy procedure performed on (b)(6) 2024.During the procedure, the cutting wire of the rx 49 broke off after stepping on the foot pedal several times during an attempt to perform a papillotomy.It was reported that no part of the cutting wire detached and fell into the patient.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.
|
| |
|
Search Alerts/Recalls
|