• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS INC. ADVIA CENTAUR CA 19-9; SYSTEM, TEST, CA19-9 FOR MONITORING AND MANAGEMENT OF PANCREATIC CANCER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS INC. ADVIA CENTAUR CA 19-9; SYSTEM, TEST, CA19-9 FOR MONITORING AND MANAGEMENT OF PANCREATIC CANCER Back to Search Results
Model Number N/A
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/14/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A customer from outside the united states reported observation of elevated ca 19-9 results with lot 535 which are discordant relative to repeat testing with lot 530.The customer indicates a shift in patient results since changing from ca 19-9 lot 530 to lot 535; the same positive shift is not observed in quality control (qc) results.Repeatability of patient-sample pools has not been a problem.The same observation is made on all three advia centaur systems in the customer¿s lab.Siemens is investigating.
 
Event or Problem Description
The customer reports observation of elevated ca 19-9 results with lot 535 which are discordant relative to repeat testing with lot 530.The customer identified a positive shift in ca 19-9 results after switching to a new lot of reagent, and re-tested patient samples using lot 530 which had been initially tested using lot 535.Lower results obtained using lot 530 were considered correct and reported to physicians.There are no known reports of patient intervention or adverse health consequences due to the elevated ca 19-9 results.
 
Additional Manufacturer Narrative
A customer from outside the united states reported observation of elevated advia centaur ca 19-9 results which were discordant relative to repeat testing.Siemens investigation confirmed an average positive bias of 56.4% with a sample population from the asia pacific region and atellica im and advia centaur 19-9 assay kit lots ending in 535 as compared to previous lots.However, investigation did not confirm the complaint allegations of commercial quality control results outside of range.Urgent medical device correction (umdc) aimc 24-14.A.Us and urgent field safety notice (ufsn) aimc 24-14.A.Ous will be distributed to customers who have received the affected product to notify them of the bias.The communication will advise customers to discontinue use of the affected product.Note: in section h6, codes for investigation findings and investigation conclusions were updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVIA CENTAUR CA 19-9
Common Device Name
SYSTEM, TEST, CA19-9 FOR MONITORING AND MANAGEMENT OF PANCREATIC CANCER
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
511 benedict ave
tarrytown NY 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
333 coney st.
registration# 1219913
east walpole MA 02032
Manufacturer Contact
barry memishian
333 coney st.
east walpole, MA 02032
5082985306
MDR Report Key19340546
Report Number1219913-2024-00313
Device Sequence Number21017687
Product Code NIG
UDI-Device Identifier00630414574028
UDI-Public00630414574028
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K031393
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 08/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/14/2024
Device Model NumberN/A
Device Catalogue Number10491244
Device Lot Number535
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/22/2024
Supplement Date Received by Manufacturer07/19/2024
Initial Report FDA Received Date05/17/2024
Supplement Report FDA Received Date08/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/14/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
1219913-07-18-2024-002-C
Patient Sequence Number1
Patient SexUnknown
-
-