• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. PRLNE BLU 2X24IN 8-0 D/A BV130-5 EP; SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. PRLNE BLU 2X24IN 8-0 D/A BV130-5 EP; SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE Back to Search Results
Catalog Number EPM8739
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information was requested, and the following was obtained: * have any of these events been previously reported to ethicon? if so, provide the respective reference number(s).>this is the first time i have reported this type of product complaint at this facility.* what are the procedure name(s) and date(s)? > procedures have been cabg (cardiac arterial bypass graft).Dates unknown.What is the quantity involved per procedure unknown, but at least once per procedure.* was there any adverse patient consequence(s) or subsequent medical/surgical intervention? > none reported.Unknown how far into each strand the needle popped off.Since this suture is typically used in a continuous baseball stitch manner, i do not know whether the needle separated at initiation or further into the closure after multiple manipulations by the surgeon.To date it has been reported that the device will not be returned.If the device or further details are received at a later date a supplemental medwatch will be sent.H6 component code: g07002 ¿ device not returned a review of the batch manufacturing records was conducted, and no related non-conformances were identified.The single complaint was reported with multiple events.There are no additional details regarding the additional patient events.Related events captured via: 2210968-2024-05409 2210968-2024-05408 2210968-2024-05407 2210968-2024-05406.
 
Event or Problem Description
It was reported that the patient underwent a coronary artery bypass graft surgery on an unknown date and suture was used.On multiple occasions, the needle "popped off" from the suture, requiring another to be opened.When asked how many times, the nurse believed there were less than 12, but could not be more specific.Two boxes of the same lot number were involved.There was no patient harm.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 8/13/2024.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRLNE BLU 2X24IN 8-0 D/A BV130-5 EP
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
calle durango #2751
ciudad juarez 32575
MX   32575
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key19341660
Report Number2210968-2024-05408
Device Sequence Number15288714
Product Code GAW
UDI-Device Identifier10705031218932
UDI-Public(01)10705031218932(11)231205(17)281130(10)TPMBST
Combination Product (Y/N)N
Initial Reporter StateNM
Initial Reporter CountryUS
PMA/510(K) Number
K133356
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberEPM8739
Device Lot NumberTPMBST
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/22/2024
Supplement Date Received by Manufacturer07/18/2024
Initial Report FDA Received Date05/17/2024
Supplement Report FDA Received Date08/13/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/05/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-