• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA USA, INC. PULSE GEN MODEL 105; GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA USA, INC. PULSE GEN MODEL 105; GENERATOR Back to Search Results
Model Number 105
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/02/2024
Type of Reportable Event Death
Additional Manufacturer Narrative
Livanova usa, inc.Submits this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation, based on information that livanova has obtained, but may not have been able to investigate or verify prior to the date the report was required by the fda.This report does not constitute an admission, or a conclusion by fda or anyone else, that the device, livanova, or livanova's employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects¿ or "malfunctions¿.These words are incorporated into the fda 3500a medwatch form by the fda, and livanova objects to their use.
 
Event or Problem Description
It was reported that the patient recently passed away.No other relevant information has been received to date.
 
Event or Problem Description
An obituary review revealed that the patient passed away peacefully in her home and the patient had been bedridden her whole life.Additional information received noting that the cause of death was hypoxic asphyxiation due to chronic asthma and no autopsy was performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 105
Common Device Name
GENERATOR
Manufacturer (Section D)
LIVANOVA USA, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
LIVANOVA USA, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
cindy scott
100 cyberonics blvd
suite 600
houston, TX 77058
2816672681
MDR Report Key19342619
Report Number1644487-2024-00571
Device Sequence Number21017710
Product Code LYJ
UDI-Device Identifier05425025750054
UDI-Public05425025750054
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P970003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date01/05/2019
Device Model Number105
Device Lot Number203996
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Event Location Other
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/24/2024
Supplement Date Received by Manufacturer05/21/2024
Initial Report FDA Received Date05/17/2024
Supplement Report FDA Received Date06/14/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/11/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient Age20 YR
Patient SexFemale
-
-