Model Number 2532
Medical Device Problem Code Device Slipped (1584)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
Customer reporting a complaint on product # 2532.Customer states the teeth are slipping & patients are able to get out of their restraints.The date the issue was discovered is unknown and no patient incident or injury was reported.
 
Additional Manufacturer Narrative
H3.The device was not returned for evaluation.Therefore, this report is based solely on the information provided by the customer.The device history record of the affected lot number 3286t099 did not reveal any issues that could have contributed to the reported incident.No ncrs were identified.The product passed all verification testing and met manufacturing specifications prior to being released for distribution.A review of the complaint database revealed similar events against this device in the past two years, where the ifu was not being followed (missing knot during set-up), thus leading tidi to redesign the male buckle of the product to mitigate the slipping issue from occurring while the knot is not present as required per the ifu.The ifu application instructions state to attach the female end of the quick-release buckle (short strap) to the frame that moves with the patient, out of the patient¿s reach (do not attach to side rail or head/footboard).You may also wrap the connecting strap once around the frame to move the buckle out of the patient¿s reach.Secure by feeding the female end through the loop in the strap.Insert the male end of the connecting strap into the female end of the short strap.Listen for a snapping sound, pull firmly on straps to ensure a good connection.To limit unwanted adjustment, tie an overhand knot with the excess strap directly below the quick-release buckle.Manufacturer reference file 2024-00489 h3 other text : product not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOAM LIMB HOLDER
Common Device Name
RESTRAINT, PROTECTIVE
Manufacturer (Section D)
TIDI PRODUCTS LLC
570 enterprise drive
neenah WI 54956
Manufacturer Contact
chris rahn
570 enterprise drive
neenah, WI 54956
9207514300
MDR Report Key19344043
Report Number2182318-2024-00065
Device Sequence Number8693005
Product Code FMQ
UDI-Device Identifier10190676004789
UDI-Public10190676004789
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number2532
Device Catalogue Number2532
Device Lot Number3286T099
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/19/2024
Initial Report FDA Received Date05/17/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/13/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-