W. L. GORE & ASSOCIATES, INC. GORE® TAG® THORACIC BRANCH ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
|
Back to Search Results |
|
| Catalog Number |
TSB121706A |
| Medical Device Problem Code |
Device Endoleak (4074)
|
| Health Effect - Clinical Code |
Clinical Endoleak (4843)
|
| Date of Event |
04/13/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Additional devices included on this report are as follows: catalog #tsb121706a/ serial #(b)(6)/ udi (b)(4).Cbas® heparin surface incorporates carmeda heparin manufactured from heparin sodium api, which is covalently bound to the device surface and is essentially non-eluting.Concomitant medical products and therapy dates: unavailable.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2024, the patient underwent endovascular treatment of an ascending aortic aneurysm in the asg22-02 arise ii pivotal study using gore® ascending stent grafts and gore® tag® thoracic branch endoprostheses (tbe).Two tbe side branch components were implanted in the patient's brachiocephalic artery.On (b)(6) 2024, a type ic endoleak was observed.A gore® excluder® aaa contralateral leg component was implanted in the patient's brachiocephalic artery and into the tbe portal.On (b)(6) 2024, the patient was discharged.There were no further reported issues.
|
| |
|
Additional Manufacturer Narrative
|
|
H.6.Investigation findings for analysis of production records: code c19 - the review of the manufacturing paperwork verified that the lots involved in this event met all pre-release specifications.H.6.Investigation conclusions: code d15 - this complaint was initiated based on information received from a clinical study database.The information reported in the complaint does not reasonably suggest that a potential device malfunction has occurred.Neither clinical images enabling direct assessment of product performance nor the product itself were returned for evaluation.Additional information was requested from the field; however, no attributable cause or contributing factors to the adverse event could be identified.There is insufficient information available for gore to reasonably draw conclusions related to aspects of the event, therefore conclusion code ¿d15: cause not established¿ is being used.Insufficient information may include limited or missing relevant medical records, involvement of multiple implanted devices (including non-gore devices) in the field of treatment, patient non-compliance, and/or a general lack of available detail or specificity related to an adverse event and/or device.H.6.Investigation conclusions: code d12 added.According to the gore® tag® thoracic branch endoprosthesis instructions for use, potential adverse events and complications that may occur with the use of the gore® tag® thoracic branch endoprosthesis include, but are not limited to, endoleak.H.6.Investigation findings for analysis of production records: code c21 updated to code c19.H.6.Investigation conclusions: code d16 updated to code d15.
|
| |
|
Search Alerts/Recalls
|
|
|