• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES LLC EDWARDS INTUITY ELITE VALVE SYSTEM; HEART-VALVE, NON-ALLOGRAFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES LLC EDWARDS INTUITY ELITE VALVE SYSTEM; HEART-VALVE, NON-ALLOGRAFT TISSUE Back to Search Results
Model Number 8300AB
Medical Device Problem Codes Microbial Contamination of Device (2303); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Endocarditis (1834); Insufficient Information (4580)
Date of Event 04/12/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was learned through implant patient registry that a 27mm 8300a aortic valve in aortic position was explanted after an implant duration of five (5) years, 11 months due to unknown cause.Avr was completed with a 29mm 11500a inspiris resilia aortic valve.The patient was in recovery post operative.Edwards lifesciences maintains an implant patient registry.This registry is a patient tracking mechanism for serialized edwards implantable devices (bioprosthetic heart valves and annuloplasty rings), rather than a true post-market surveillance registry.Through the registry, edwards is notified when these devices are implanted.In addition, patient and/or device status may be reported to the registry via the implantation data cards.The information is received from various sources (e.G.Surgeon, hospital, and patient family members) and is not received in the form of a conventional "customer complaint".The information reported may or may not be related to the edwards device.
 
Additional Manufacturer Narrative
H11: additional manufacturer narrative: the investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Additional Manufacturer Narrative
H11: additional manufacturer narrative: updated sections: b4, b5, b7, g3, g6, h2, h6 (clinical code, device code, and type of investigation).H11: corrective data: based on the additional information obtained, this event is no longer considered reportable, and this correction is being submitted.
 
Event or Problem Description
It was learned through implant patient registry and investigation that a 27mm 8300a aortic valve in aortic position was explanted after an implant duration of five (5) years, 11 months due to mssa endocarditis and bacteremia with severe aortic stenosis.The patient presented with fevers, chills, diaphoresis, lethargy, and malaise.Avr was completed with a 29mm 11500a inspiris resilia aortic valve.The patient was in recovery post operative.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS INTUITY ELITE VALVE SYSTEM
Common Device Name
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
EDWARDS LIFESCIENCES LLC
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES LLC
1 edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
1 edwards way
mle fl2- office m2013
irvine, CA 92614
9492506615
MDR Report Key19347234
Report Number2015691-2024-03723
Device Sequence Number6614728
Product Code LWR
UDI-Device Identifier00690103194517
UDI-Public(01)00690103194517(17)220108
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P150036
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/08/2022
Device Model Number8300AB
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/23/2024
Supplement Date Received by Manufacturer05/22/2024
Initial Report FDA Received Date05/17/2024
Supplement Report FDA Received Date06/17/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/09/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Life Threatening; Required Intervention;
Patient Age73 YR
Patient SexMale
-
-