• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH V40 COCR LFIT HEAD 40MM/-4; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH V40 COCR LFIT HEAD 40MM/-4; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Catalog Number 6260-9-040
Medical Device Problem Codes Material Erosion (1214); Patient Device Interaction Problem (4001)
Health Effect - Clinical Codes Cyst(s) (1800); Metal Related Pathology (4530)
Date of Event 04/26/2024
Type of Reportable Event Serious Injury
Event or Problem Description
Post tha surgery, there was a finding of bone atrophy in the anterior of the acetabulum, and armd (adverse reaction to metal debris) was suspected.
 
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Review of the device history records indicate devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.H3 other text : not returned.
 
Additional Manufacturer Narrative
An event regarding other (cystic lesion) involving a metal head was reported.The event was confirmed via clinician review of the provided medical records.Method & results: -product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device remains implanted.-clinician review: a review of the provided medical records by a clinical consultant indicated: "the x-rays reviewed demonstrate a large cystic lesion lateral to the acetabular component.A peri-acetabular cystic lesion is confirmed.The root cause of this cystic lesion cannot be determined from the documentation provided." -product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient is planned to be revised due to bone atrophy in the anterior of the acetabulum and suspected armd (adverse reaction to metal debris).A review of the provided medical records by a clinical consultant indicated: "the x-rays reviewed demonstrate a large cystic lesion lateral to the acetabular component.A peri-acetabular cystic lesion is confirmed.The root cause of this cystic lesion cannot be determined from the documentation provided." the periprosthetic cyst is confirmed but cannot be attributed to a specific device based on the information provided.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
Post tha surgery, there was a finding of bone atrophy in the anterior of the acetabulum, and armd (adverse reaction to metal debris) was suspected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V40 COCR LFIT HEAD 40MM/-4
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
brad curtis
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key19348334
Report Number0002249697-2024-00750
Device Sequence Number17928945
Product Code JDI
UDI-Device Identifier07613327032703
UDI-Public07613327032703
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K061434
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2019
Device Catalogue Number6260-9-040
Device Lot NumberMNR709
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/26/2024
Supplement Date Received by Manufacturer07/31/2024
Initial Report FDA Received Date05/18/2024
Supplement Report FDA Received Date08/26/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/10/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age63 YR
Patient SexFemale
Patient Weight55 KG
Patient RaceAsian
-
-