| Catalog Number |
8065993245 |
| Medical Device Problem Code |
Dull, Blunt (2407)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
04/22/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A healthcare professional reported that four of the ophthalmic blades were found to be dull.Procedure details and patient impact were not reported.Additional information has been requested but none received till date.
|
| |
|
Additional Manufacturer Narrative
|
|
A sample was not received at the manufacturing site for evaluation for the report of the blade is very blunt; therefore, the condition of the product could not be verified.A review of the device history record traceable to the reported lot numbers, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.A sample was not received at the manufacturing site and the device history record review of the lot number provided indicated the product was processed and released according to the product¿s acceptable criteria, therefore; the root cause for the customer complaint issue cannot be determined.The exact root cause for this complaint is unknown, therefore, specific action with regards to this complaint cannot be taken.All knives are 100% inspected by trained operators using a minimum of 10x magnification during manufacturing.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No additional action is required at this time.The manufacturer internal reference number is: (b)(4),.
|
| |
|
Search Alerts/Recalls
|