|
Investigation summary: one (1) sample was returned under the p/n: di-60hl, l/n: 4305708 for evaluation with its original packaging, in decontaminated conditions.The complaint sample was visually inspected, at 12¿ to 16¿ under normal conditions of illumination.Missing parts (cap, spike, vented 00-106) in the sample were detected.The sample was assembled in the level 1 system equipment to introduce distillated water and verify the correct functionality of the sample.During the test, a leak was found in union of tube and end cap, the failure mode "a0282 - leaking" reported in the complaint was confirmed.However, a root cause was not determined.A review of the device history record (dhr) shows there were no observations or nonconformities recorded during manufacture to suggest an issue of this nature would occur with this lot of products.
|