| Model Number |
SCORPION NEEDLE, KNEE |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.The complaint is confirmed.(1) unpackaged ar-12990 was received with a broken needle, ar-12990n-1, for evaluation.Visual inspection noted the needle was broken and stuck inside the cannula of the ar-12990.The broken fragments were not returned.The most likely cause for the reported failure can be attributed to user error of the device due to excessive force being used during firing of the needle and/or unsecure grasp of tissue.
|
| |
|
Event or Problem Description
|
|
It was reported that during a knee arthroscopy the needle broke inside the knee scorpion.According to the surgeon there was no harm for patient, operator or third party reported.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|