| Catalog Number |
02-012-35-2011 |
| Medical Device Problem Codes |
Unstable (1667); Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Scar Tissue (2060); Synovitis (2094); Osteolysis (2377); Joint Laxity (4526); Implant Pain (4561); Joint effusion (4849)
|
| Date of Event |
11/17/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that approximately 167 months after a right total knee replacement procedure, the patient underwent a revision procedure to address prosthesis wear, instability, bone loss, and osteolysis.Revision surgery operative notes were provided.Patient left the operating room in stable condition.Post operative diagnosis noted wear of the implant.No further issues or complications were reported.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
D10: concomitant devices are unknown.The device was not returned for evaluation and no photographs or x-rays were provided for review; therefore the reported event cannot be confirmed through analysis.The cause of the reported event cannot be determined based on the information made available.Should additional, relevant information become available, a supplemental report will be filed accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
The reason for the revision reported in (b)(4) cannot be confirmed from the information provided but may be the result of prosthesis wear, instability, and femoral loosening or due to inclusion of the polyethylene in the packaging recall.The reported femoral loosening may have been the result of failure of the cement used to secure the femoral component to the femur, which led to loosening at the cement-implant interface.Additionally, these devices appear to have been implanted for over ten years prior to the reported event.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.H6: corrected health effect and medical device clinical codes.
|
| |
|