| Model Number |
20001100100011360 |
| Medical Device Problem Code |
Failure of Device to Self-Test (2937)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during functional testing, the device prompted a "unit failed" message.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental medwatch reports the evaluation of the device.This supplemental medwatch report also corrects information submitted in the initial medwatch report.Please reference sections d4 model number updated, d4 catalog number removed, d4 primary udi number updated.Evaluation results: the device was evaluated by zoll medical corporation.The customer's report was not replicated or confirmed.The device was put through extensive testing which included bench handling, power cycling, and functional stress testing using the returned pads without duplicating the customer's report.The device was replaced and returned to the customer.No trend is associated with reports of this type.
|
| |
|