• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CURONIX LLC FREEDOM PERIPHERAL NERVE STIMULATOR; TRANSMITTER ASSEMBLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CURONIX LLC FREEDOM PERIPHERAL NERVE STIMULATOR; TRANSMITTER ASSEMBLY Back to Search Results
Model Number PDBT-915-2K
Medical Device Problem Code Battery Problem (2885)
Health Effect - Clinical Code Burning Sensation (2146)
Date of Event 05/07/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
As a result of the loose battery, the battery was punctured and leaked, and cannot delivery therapy.For the report of overheating: thermal imaging was used to verify the outside temperature of the device while charging, the internal temperature of the device while charging, the outside temperature of the device while operating, and the internal temperature of the device while operating.Results are as follows: outside thermal temperature while charging: 30.5°c.Outside thermal temperature while operating: 29.4°c.Internal thermal temperature while operating: 31.0°c.The outside thermal temperature while charging was 30.5°c, which is below the product requirements specification of 41°c (106°f).Based on the thermal imaging, there is no evidence of thermal damage and the product met temperature specifications.Reference capa-0042 for additional investigation details for the loose battery.
 
Event or Problem Description
The patient reported their transmitter assembly was getting hot and caused a burning sensation to the skin.The overheating did not cause tissue damage or require medical intervention.The patient was provided a replacement transmitter assembly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREEDOM PERIPHERAL NERVE STIMULATOR
Common Device Name
TRANSMITTER ASSEMBLY
Manufacturer (Section D)
CURONIX LLC
1310 park central blvd s
pompano beach FL 33064
Manufacturer (Section G)
CURONIX LLC
1310 park central blvd s
pompano beach FL 33064
Manufacturer Contact
tena jimmerson
1310 park central boulevard s.
pompano beach, FL 33064
8009655134
MDR Report Key19354974
Report Number3010676138-2024-00090
Device Sequence Number7678051
Product Code GZF
UDI-Device Identifier00818225020150
UDI-Public(01)00818225020150
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K171366
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberPDBT-915-2K
Device Lot Number00372-S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/14/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/07/2024
Initial Report FDA Received Date05/20/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/13/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-