| Model Number |
EPK-I7010 |
| Medical Device Problem Code |
Product Quality Problem (1506)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/09/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The time of event is not during procedure.There was no report of patient harm.Iris diaphragm failure.
|
| |
|
Additional Manufacturer Narrative
|
|
This device is classified as import for export, therefore 510k is not applicable.Model epk-i7010-us is available in the usa with a 510k number k182004.The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the iris diaphragm as defective.Based on the result, we concluded that it was caused due to the excessive shock applied on the iris diaphragm.Based on the technical report ""hr-rpt-0973(temperature rise)"" and/or the risk analysis results, it was evaluated to submit mdr.
|
| |
|