MAUDE Adverse Event Report: TELEFLEX MEDICAL ET TUBE, HVT, 8.0; TUBE, TRACHEAL (W/WO CONNECTOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL ET TUBE, HVT, 8.0; TUBE, TRACHEAL (W/WO CONNECTOR Back to Search Results
Catalog Number 5-10316
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Codes Hypoxia (1918); Insufficient Information (4580)
Date of Event 04/13/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Qn# (b)(4).The customer returned one unit 5-10316 et tube,hvt,8.0 for investigation.The et tube was visually examined with and without magnification.Visual examination of the returned device revealed that the sample appeared typical.No defects or anomalies were observed.Functional inspection was performed on the returned device to test for proper inflation and deflation.A lab inventory syringe was filled with air and attached to the pilot balloon.Upon injection of air, the balloon cuff was unable to stay inflated.The device was submerged underwater in order to test for leaks.Upon inflation , the device leaked from the balloon cuff.Multiple holes were found in a line on the balloon cuff.It appeared that the cuff was cut by something sharp.The ifu for this product states, "the tracheal tube cuff inflation system, including the valve, should be checked prior to intubation.If a malfunction is a detected in any part of the system, the tube should not be used." "care should be taken to avoid damaging the thin-walled cuffs during intubation.If cuff is damaged, the tube should not be used." the ifu also states, "cuff pressure should be monitored routinely to assure that an adequate tracheal seal is maintained and that the cuff has not become over-inflated.The tracheal tube cuff should be slowly inflated with the minimum amount of air required to provide an effective tracheal seal.Do not inflate with a measured volume of air or by "feel" of pressure in the syringe since little resistance should be felt during inflation." the reported complaint was confirmed based upon the sample received.The returned et tube was found to have multiple holes in a line in the balloon cuff which prevented it from being able to stay inflated.All et tubes are 100% inspected for inflation and deflation at the time of manufacturing.It is unlikely that this type of damage was present at the time of manufacturing.Therefore, based upon the damage observed, unintentional user error caused or contributed to this event.
 
Event or Problem Description
It was reported that: "pilot balloon ett failure.Initial failure may be related to agitation and possibly patient biting tube." associated complaints 3003898360-2024-00494, 3003898360-2024-00493 and 3003898360-2024-00495.
 
Additional Manufacturer Narrative
Qn#(b)(4).The customer returned one unit 5-10316 et tube,hvt,8.0 for investigation.The et tube was visually examined with and without magnification.Visual examination of the returned device revealed that the sample appeared typical.No defects or anomalies were observed.Functional inspection was performed on the returned device to test for proper inflation and deflation.A lab inventory syringe was filled with air and attached to the pilot balloon.Upon injection of air, the balloon cuff was unable to stay inflated.The device was submerged underwater in order to test for leaks.Upon inflation , the device leaked from the balloon cuff.Multiple holes were found in a line on the balloon cuff.It appeared that the cuff was cut by something sharp.The ifu for this product states, "the tracheal tube cuff inflation system, including the valve, should be checked prior to intubation.If a malfunction is a detected in any part of the system, the tube should not be used." "care should be taken to avoid damaging the thin-walled cuffs during intubation.If cuff is damaged, the tube should not be used." the ifu also states, "cuff pressure should be monitored routinely to assure that an adequate tracheal seal is maintained and that the cuff has not become over-inflated.The tracheal tube cuff should be slowly inflated with the minimum amount of air required to provide an effective tracheal seal.Do not inflate with a measured volume of air or by "feel" of pressure in the syringe since little resistance should be felt during inflation." the reported complaint was confirmed based upon the sample received.The returned et tube was found to have multiple holes in a line in the balloon cuff which prevented it from being able to stay inflated.All et tubes are 100% inspected for inflation and deflation at the time of manufacturing.It is unlikely that this type of damage was present at the time of manufacturing.Therefore, based upon the damage observed, unintentional user error caused or contributed to this event.Corrected data: uppdate to imdrf codes and incident description.
 
Event or Problem Description
It was reported that: "pilot balloon ett failure.Initial failure may be related to agitation and possibly patient biting tube.Patient was hypoxemic and not getting appropriate volumes.Required prolonged bagging." associated complaints 3003898360-2024-00494, 3003898360-2024-00496, 3003898360-2024-00493 and 3003898360-2024-00495.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ET TUBE, HVT, 8.0
Common Device Name
TUBE, TRACHEAL (W/WO CONNECTOR
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
rancho el descanso
tecate 21478
MX   21478
Manufacturer Contact
elaine cully
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key19357707
Report Number3003898360-2024-00496
Device Sequence Number12366008
Product Code BTR
UDI-Device Identifier04026704397051
UDI-Public04026704397051
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K822082
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number5-10316
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/26/2024
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/25/2024
Supplement Date Received by Manufacturer05/14/2024
Initial Report FDA Received Date05/21/2024
Supplement Report FDA Received Date05/21/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NOT REPORTED; NOT REPORTED
Patient SexUnknown
-
-