The device was not returned for evaluation.Production record and corrective and preventive actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a review of the complaint handling database identified no similar complaints from this lot.A conclusive cause for the reported material rupture could not be determined; however, factors that may contribute to material ruptures include, but are not limited to, material damage, inflation technique, interaction with accessory devices, lesion calcification and tortuosity or insufficient preparation prior to use.There was no damage noted to the stent delivery system (sds) during the inspection prior to use which suggests a product quality issue did not contribute to the reported material rupture.In this case, it is possible the device may have interacted with the heavily calcified, moderately tortuous, 90% stenosed lesion during advancement/positioning, as resistance was noted, causing the reported material rupture; however, without the device to examine, this cannot be confirmed.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
|