• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. IBAL UKA TIB BEA-RING TRL SZ 1 9MM; ORTHOPEDIC MANUAL SURG INSTR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. IBAL UKA TIB BEA-RING TRL SZ 1 9MM; ORTHOPEDIC MANUAL SURG INSTR Back to Search Results
Model Number IBAL UKA TIB BEA-RING TRL SZ 1 9MM
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 09/23/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.Since the complaint device was not returned, a physical evaluation of the device was not performed.However, the reported event is not confirmed without the device being returned or any photos provided.The most likely cause for the reported failure can be attributed to user error of the device due to prying/leveraging the device during use.
 
Event or Problem Description
On 9/23/2021, it was reported by a sales representative via (b)(4) that an ar-601-tba8 tibial bearing trial 8mm and ar-601-tba9 tibial bearing trial 9mm broke.This was discovered during a procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IBAL UKA TIB BEA-RING TRL SZ 1 9MM
Common Device Name
ORTHOPEDIC MANUAL SURG INSTR
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key19359742
Report Number1220246-2024-03590
Device Sequence Number17746270
Product Code HWT
UDI-Device Identifier00888867037458
UDI-Public00888867037458
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIBAL UKA TIB BEA-RING TRL SZ 1 9MM
Device Catalogue NumberAR-601-TBA9
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/23/2021
Initial Report FDA Received Date05/21/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-