| Model Number |
3334800 |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/22/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during use, the visao drill handpiece was getting overheated.The procedure was completed using another drill handpiece.There was no patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: analysis results were not available as of the date of this report.A follow-up report will be submitted when analysis is complete.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Pre-repair temperature test of the device performed for 1 min, found that at nose - 114.9f, middle - 115.9f and rear - 84.4f which have not exceeded 118.4 f hence does not meet the reporting criteria.
|
| |
|
Additional Manufacturer Narrative
|
|
As per the pre-repair temperature test of the device for 1 min, it was found that at nose - 114.9f, middle - 115.9f and rear - 84.4f which were less than 118.4 f hence this does not meet the reporting criteria.Above information received shows that there is no information to reasonably suggest that the device in this report may have caused or contributed to a death or serious injury or that the device in this report has malfunctioned.Therefore, this event does not meet the reporting requirements in 21 cfr 803.Correction: g2: report source of "foreign" was not check marked in previous report.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|