This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, d2, g1-2, g3, g7, h1, h2, h6, h10.It is unknown which screw fractured.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Device history record was reviewed and no discrepancies were found.The root cause of the reported issue is attributed to user error as the patient was lifting a hot water heater post-op.Per ifu, "excessive, unusual and/or awkward movement and/or activity, trauma, excessive weight, and obesity have been implicated with premature failure of certain implants by loosening, fracture, dislocation, subluxation and/or wear." the reported event is confirmed as the patient was lifting a hot water heater post-op causing the screw to fracture.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|