Model Number SIF-Q180
Medical Device Problem Code Failure to Clean Adequately (4048)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/07/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was observed during the device inspection, that the small intestinal videoscope exhibited white foreign material inside the air/water tube and air/water cylinder.There were no reports of patient involvement.
 
Additional Manufacturer Narrative
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Additional Manufacturer Narrative
This supplemental report is to inform that upon further review, this is not a reportable malfunction.The initial medwatch reported that during the device inspection, the small intestinal videoscope exhibited white foreign material inside the air/water tube and air/water cylinder.However, the device was returned without reprocessing due to other (non-reportable) defects, which caused the foreign material to remain.Per the legal manufacturer, this issue is not likely to cause or contribute to death or serious injury if the malfunction were to recur.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II SMALL INTESTINAL VIDEOSCOPE
Common Device Name
SMALL INTESTINAL VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key19362704
Report Number9610595-2024-10262
Device Sequence Number15038577
Product Code FDA
UDI-Device Identifier04953170339936
UDI-Public04953170339936
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K071254
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberSIF-Q180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/14/2024
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/17/2024
Supplement Date Received by Manufacturer06/01/2024
Initial Report FDA Received Date05/21/2024
Supplement Report FDA Received Date06/12/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1