|
Product analysis #706383084: equipment used: video inspection system (m-78210), ruler 200cm (m-83361); as found condition: the pipeline vantage shield embolization device and phenom 27 catheter were returned for analysis within a shipping box; within a plastic bio-pouch; within an opened pipeline vantage shield outer carton; and within a plastic bag.The pipeline vantage shield pusher was returned outside the phenom 27 catheter.The pipeline vantage shield braid was not returned for analysis.Therefore, any contributing factors could not be assessed.Damage location details: no bent or kink was found with pushwire.However, the pushwire was found to be separated/broken proximal to the dps sleeves.The tip coil and dps sleeves were not returned for analysis and the reason was not provided.No damages were found with arms.The distal hypotube and ptfe shrink tubing were found to be intact with no signs of elongation.The distal and proximal dps restraints were found to be intact.The dps sleeves were found intact with no signs of damage.No flash or voids molded were observed in the hub.No damage was found with hub.The catheter body was found to be kinked at ~23.2cm from proximal end.The phenom 27 catheter tip and marker were examined; and no damages were found.No other anomalies were observed.Testing/analysis: the total and usable lengths of phenom 27 catheter were measured to be within specifications.The catheter was flushed with water and water exited out of the distal tip.The catheter was then tested by running an in-house 0.0265¿ mandrel through catheter tip and hub.The mandrel successfully passed through the catheter hub and tip with no issues; however, the resistance observed at the damaged locations.The broken end was sent out for sem (scanning electron microscopy) analysis.Conclusion: based on the analysis findings, the pipeline vantage shield and phenom 27 catheter were confirmed to have resistance during delivery.The phenoom 27 was found to be damaged.From the damages seen on the catheter body (kinking); it is likely high force used during the delivery.Additionally, the pushwire was found to be separated proximal to the dps sleeves.Per the sem result, the broken end failed via mechanical damage (cutting/shearing).It is likely these damages occurred when the customer attempted to advance and retrieve the pipeline vantage shield through the phenom 27 catheter against resistance.The review of lot history records shows that the finished device has met all manufacturing requirements and specifications during final assembly and quality inspection.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|