| Model Number |
9735665 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/29/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: 9735740, version: 2.1.0.H3: a medtronic representative went to the site to test the equipment.Testing revealed that no failures were found.The navigation system then passed the system checkout and was found to be fully functional.H6: fdm b01, fdr c19, fdc d14 are applicable to the system checkout.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a navigation system being used during a sacroiliac and thoracolumbar procedure.It was reported that the site experienced an alleged inaccuracy while in surgery.They had previously taken a spin that they were confident in but after taking a second, the surgeon felt off by an unknown distance and direction.They took a third spin in attempt to resolve but still felt inaccurate.There was a reported delay to the procedure of 2 hours due to this issue.There was no reported impact on patient outcome.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: a software analysis was initiated.However, it was found that there was insufficient information to determine the relationship of the software to the reported issue.H6: fdm b01, fdr c19, and fdc d15 are applicable to the software analysis.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|