| Model Number |
IBALANCE TKA, HANDLE, MODULAR KEEL PUNCH |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/19/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the ar-613-1 is cracked.The issue was discovered upon inspection.
|
| |
|
Additional Manufacturer Narrative
|
|
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.Complaint confirmed.Visual evaluation confirms the handle is cracked at distal end.The cause is undetermined, however the most likely cause for the reported failure can be attributed to user-applied excessive mechanical forces during use.
|
| |
|