Model Number 8282WL
Medical Device Problem Code No Audible Alarm (1019)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H3 investigation into the returned products did not confirm the reported issues.Visual findings observed the fall monitor and wireless nurse call adapter both being received with batteries inside.Product was not returned for 8309wl as it was discarded by customer.The returned fall monitor unit was evaluated with the returned wireless nurse call adapter and an in-house wireless sensor pad.The unit was tested by simulating use.It functioned as expected, passing all functional tests.No physical damage was observed on either the fall monitor nor the wireless nurse call adapter.Instructions for use (ifu) were reviewed and found to provide adequate instructions and warnings for safe and effective use of the device.The ifu states if the alarm and/or sensor do not function properly, remove the alarm and sensor from service and replace them with a properly functioning alarm and/or sensor.Do not use the alarm or sensor if it does not activate each time weight is removed from the sensor, or if the chair belt sensor is unfastened.At this time there is no evidence that a manufacturing non-conformity contributed to the reported complaint.All complaints are trended and reviewed by management on a monthly basis.As a part of the monthly review, any excursion above the control limits for this failure mode will be assessed, documented and acted upon as warranted.Manufacturer reference file (b)(4).
 
Event or Problem Description
Customer reporting a complaint on product #'s 8645wl, 8282wl, & 8309wl.Customer states that a fall occurred with no alarm sounding.When the staff tested the alarm they noticed that it would pair and then detach without sounding an alarm or voice/noise sounding.The pad was not kept & was discarded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WIRELESS NURSE CALL ADAPTER
Common Device Name
MONITOR, BED PATIENT
Manufacturer (Section D)
TIDI PRODUCTS LLC
570 enterprise drive
neenah WI 54956
Manufacturer Contact
chris rahn
570 enterprise drive
neenah, WI 54956
9207514300
MDR Report Key19370586
Report Number2182318-2024-00067
Device Sequence Number7703227
Product Code KMI
UDI-Device Identifier10190676007643
UDI-Public10190676007643
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number8282WL
Device Catalogue Number8282WL
Device Lot Number2249KQ0195
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/06/2024
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/26/2024
Initial Report FDA Received Date05/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/28/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-