EXACTECH, INC. NV GXL LNR, NEUTRAL, 28MM ID, GROUP 2 CUPS; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
|
Back to Search Results |
|
| Catalog Number |
130-28-52 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Cyst(s) (1800); Failure of Implant (1924); Scar Tissue (2060); Osteolysis (2377); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
03/19/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
D10: concomitant devices: 2054905 120-65-35 - bone screw 6.5mm dia x 35mm long 2134091 142-28-07 - cocr fem head 28mm +7 offset 12/14 2313339 118-41-01 - p-series pf plasma h/o collarless sz 1 12/14 2454270 101-05-01 - drill bit 1 pk 3.2mm dia x 32mm lng 2491586 180-01-52 - nv crown cup clstr hole 52mm group 2 2606123 120-65-15 - bone screw 6.5mm dia x 15mm long 2684492 120-65-15 - bone screw 6.5mm dia x 15mm long.The product associated with the reported event is within the scope of recall z-1729-2022; however, there is insufficient information to evaluate whether the subject issue of the recall was the cause or contributor to the reported event.The device was not returned for evaluation and no medical or other records containing treatment information or patient information have been received; therefore, the reported event cannot be confirmed, nor can the circumstances or potential causes or contributors to the alleged event be evaluated.Should additional material information become available that permits more analysis or conclusions, a supplemental report will be filed accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported that approximately 130 months after a right total hip replacement procedure, the patient underwent a revision procedure to address prosthesis wear, worsening osteolysis, and cyst superior to acetabular component.Revision surgery operative notes were provided.Patient left the operating room in stable condition.Post operative diagnosis noted wear of the implant.No further issues or complications were reported.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The most likely underlying cause for the revision reported is a combination of risk factors specified such as use error, implant positioning, implant size selection, and patient factors (fitness for surgery, biomechanics, activity level and local tissue oxidation potential) and being implanted for more than 10 years.However, this cannot be confirmed from the reported information and the devices were not available for evaluation.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.".
|
| |
|