| Model Number |
39347-070470 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/26/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D2b: product code: dqy device evaluated by mfr: device was returned for analysis.A visual examination identified that the balloon was not folded which indicates that the device was subjected to positive pressure.A leak test identified a balloon pinhole located approximately 1mm distal of the distal markerband.As per specification, the rated burst pressure for this device is 34 atmospheres.A visual examination observed no damage to the tip of the device.A visual and tactile examination found no kinks or damage to the shaft of the device.Both markerbands were undamaged and present on the shaft of the device.
|
| |
|
Event or Problem Description
|
|
Reportable based on the device analysis completed on (b)(6) 2024.It was reported that balloon leakage occurred.The target lesion was located in the arteriovenous fistula.A 7.0mm x 40mm x 75cm athletis balloon catheter was used for dilatation.During the procedure, it was noted that the contrast medium had leaked out.The procedure was completed with a different device.There were no patient complications reported.However, device analysis revealed a balloon pinhole.
|
| |
|
Search Alerts/Recalls
|