• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER ORTHOPAEDIC MFG. LTD. PERSONA ARTICULAR SURFACE MEDIAL CONGRUENT LEFT 12 MM; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER ORTHOPAEDIC MFG. LTD. PERSONA ARTICULAR SURFACE MEDIAL CONGRUENT LEFT 12 MM; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Adhesion(s) (1695); Pain (1994); Loss of Range of Motion (2032); Scar Tissue (2060); Swelling/ Edema (4577)
Date of Event 04/25/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the device remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Additional associated products(reported): 42508606401, persona cruciate retaining left size 8 pps std femur, lot #: 66126127.Additional associated products(not reported): 42535007501, 0 spiked keel left size f osseoti tibia, lot #: 66273434.42540200035, all-poly patella cemented 35 mm diameter, lot #: 66041659.110034355, refobacin bc r 1x40 us, lot #: unk.
 
Event or Problem Description
It was reported a patient began to experience limited range of motion and stiffness about one month post implantation.Patient underwent a manipulation under anesthesia about one week later with no complications.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported a patient began to experience limited range of motion and stiffness about one month post implantation.Patient underwent a manipulation under anesthesia about one week later with no complications.Following the manipulation, the patient continued to experience stiffness and limited range of motion and two weeks after the first mua a second manipulation was completed.All implants remain in place and the event is improving.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Arthrofibrosis is defined as the development of fibrous scar tissue within or surrounding a joint.Arthrofibrosis is a known postoperative procedure related complication that can occur from surgical implantation of new joint replacement as well as from previous injuries or surgical procedures.Scar tissue formation is a normal healing response; however, the buildup of such can result in pain, stiffness, limited range of motion, and difficulty properly ambulating.If excess scar tissue develops, conservative measures such as exercises or physical therapy would be attempted first.If these attempts fail, surgical intervention such as manipulation under anesthesia, arthroscopic arthrolysis, or open arthrotomy would become necessary to remove the fibrous tissue and restore joint function.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: b4, b5, g3, g6, h1, h2, h6, h11.An updated investigation is in process.Once the updated investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported a patient began to experience limited range of motion and stiffness about one month post implantation.Patient underwent a manipulation under anesthesia about one week later with no complications.Following the manipulation, the patient continued to experience stiffness and limited range of motion and two weeks after the first mua a second manipulation was completed.In the months following, the patient presented with continued stiffness.Five months after the second manipulation, the patient underwent an arthroscopic lysis of adhesions and synovectomy due to arthrofibrosis.All implants remain in place and the event is improving.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Arthrofibrosis is defined as the development of fibrous scar tissue within or surrounding a joint.Arthrofibrosis is a known postoperative procedure related complication that can occur from surgical implantation of new joint replacement as well as from previous injuries or surgical procedures.Scar tissue formation is a normal healing response; however, the buildup of such can result in pain, stiffness, limited range of motion, and difficulty properly ambulating.If excess scar tissue develops, conservative measures such as exercises or physical therapy would be attempted first.If these attempts fail, surgical intervention such as manipulation under anesthesia, arthroscopic arthrolysis, or open arthrotomy would become necessary to remove the fibrous tissue and restore joint function.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Event or Problem Description
It was reported a patient began to experience limited range of motion and stiffness about one month post implantation.Patient underwent a manipulation under anesthesia about one week later with no complications.Following the manipulation, the patient continued to experience stiffness and limited range of motion and two weeks after the first mua a second manipulation was completed.In the months following, the patient presented with continued stiffness.Five months after the second manipulation, the patient underwent an arthroscopic lysis of adhesions and synovectomy due to arthrofibrosis.All implants remain in place and the event is improving.Suspected tibial overhang is noted in supplied records.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.A new updated investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.D10: additional associated products 42508606401 persona cruciate retaining left size 8 pps std femur lot# 66126127.42535007501 persona 0 deg spiked keel left size f osseoti tibia lot# 66273434.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected.Updated: b4, b5, d5, g3, g6, h1, h2, h3, h6, h10, h11.D10: 42512100812 persona articular surface medial congruent left 12 mm lot#: 65752458.42508606401 persona cruciate retaining left size 8 pps std femur lot#: 66126127.42535007501 persona 0 deg spiked keel left size f osseoti tibia lot#: 66273434.42540200035 persona all-poly patella cemented 35 mm dia lot#: 66041659.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing.The reported products were reviewed for compatibility with no issues noted.Medical records were provided and reviewed by a health care professional.Review of the available records identified the following: initial; mua, preop rom 5-90, mua performed without complications with rom 3-1258; mua continued stiffness and limited rom 80 degrees following mua.A second mua was completed to resolve.; arthroscopic arthrolysis: continued stiffness following mua and pt, diagnosed with left knee arthrofibrosis 4 months after mua.Left knee scope with synovectomy and lysis of adhesions on 5 months post mua, no implants exchanged.4 months later5: persistent left knee pain, intermittent swelling, concern of overhang, aspiration of fluid from left moderate knee effusion, cortisone injection in left knee, event resolving, patient declined study continuation due to ongoing left knee pain.The reported event was confirmed via medical records.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERSONA ARTICULAR SURFACE MEDIAL CONGRUENT LEFT 12 MM
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
Manufacturer (Section G)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI  
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key19373880
Report Number3007963827-2024-00195
Device Sequence Number15384716
Product Code JWH
UDI-Device Identifier00889024469310
UDI-Public(01)00889024469310(17)280406(10)65752458
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K150090
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Study,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 09/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberN/A
Device Catalogue Number42512100812
Device Lot Number65752458
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/30/2024
Supplement Date Received by Manufacturer06/03/2024
10/10/2024
10/29/2024
03/11/2025
04/10/2025
08/27/2025
Initial Report FDA Received Date05/22/2024
Supplement Report FDA Received Date06/26/2024
10/11/2024
11/26/2024
03/14/2025
05/08/2025
09/19/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/24/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11 NARRATIVE.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexMale
Patient Weight84 KG
Patient EthnicityNon Hispanic
Patient RaceBlack Or African American
-
-