Product complaint #(b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: d2b.Additional pro-code: jey.D9: complainant part is not expected to be returned for manufacturer review/investigation.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
It was reported that the complaint regarding the 6mm self drilling matrix midface screws.While doing a midface case with one of the surgeons and four of the 6mm screws broke while being inserted into the patient's zf suture.The surgeon tried to remove them to reposition the plate but all four heads broke off.He attempted to remove them but was unable to.He ended up burring them down, repositioned the plate and moved on.Another set (matrix combo) was pulled and utilized the screws in this set which worked flawlessly.Surgery completed successfully, patient status/ outcome in good standing.This report is for one (1) ti matrixmidface emergency screw 6mm.This is report 2 of 4 for complaint (b)(4).
|