| Model Number |
N/A |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/30/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A customer from outside the united states reported observation of elevated atellica im enhanced estradiol (ee2) results which were discordant relative to repeat testing.Siemens is investigating.
|
| |
|
Event or Problem Description
|
|
The customer reports observation of an elevated atellica im enhanced estradiol (ee2) result which was discordant relative to repeat testing.The customer observed ee2 quality control (qc) results above acceptance ranges, and repeat-tested samples which had been tested when qc was out.The initial elevated result was reported to the physician.The sample was repeat-tested and a lower result was obtained.According to the customer, no corrected report was issued since the initial result was not questioned by the physician.There are no known reports of patient intervention or adverse health consequences due to the elevated ee2 result.
|
| |
|
Additional Manufacturer Narrative
|
|
Mdr was initially filed on 22-may-2024.Additional information (25-jun-2024): siemens healthcare diagnostics has concluded the investigation of the event.A customer from outside the united states reported observation of elevated quality control (qc) and patient results when using atellica im enhanced estradiol (ee2), lot 335.Siemens evaluated the instrument data and the information provided by the customer.The customer performed a recalibration of the ee2 assay.After recalibration, the quality control (qc) results recovered within the laboratory expected ranges.Service was dispatched to the customer site and the instrument¿s event log was reviewed.Close review of the instrument data showed the primary and secondary vacuum were running high.Siemens customer service engineer (cse) performed the routine vacuum system troubleshooting.Following this routine troubleshooting intervention, qc results were stabilized, and precision was acceptable.The issue appears to be resolved although a specific cause for this event was not identified.A product problem has not been identified.The customer is operational and no further actions are required.Note: in section h6, codes for investigation findings and investigation conclusions were updated.
|
| |
|
Search Alerts/Recalls
|