Section a2, a3, a4 and a5: per regulation eu 2016/679 (general data protection regulation), patient identifiers were not collected or recorded and therefore are not available.Section b3 - date of event: unknown, as information was requested but not provided.Section d6a - implant date: not applicable, as there is no indication that the lens was implanted.Section d6b - explant date: not applicable, as there is no indication that the lens was implanted.Section e1 - telephone number: (b)(6).Device evaluation: product testing could not be performed because the device was not returned.Therefore, the reported issue could not be verified, and product quality deficiency could not be determined.Manufacturing record review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.A search of complaints related to this production order (po) was performed.The search revealed that no additional complaints were received for this po.Therefore, no escalation required.Conclusion: as a result of the investigation there is no indication of a product quality deficiency, and the reported issue could not be verified.Attempts have been made to obtain the missing information.However, to date, no response has been received.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|