| Model Number |
LO-PRO LOCK SCRW,SS 3.5X 14MM |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
11/12/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device has loosened in the clavicle plate.There was no harm for patient, operator or third party reported.This was noticed after the surgery.No further information received.
|
| |
|
Additional Manufacturer Narrative
|
|
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The most likely cause for the reported failure can be attributed to user error due to misaligned insertion and/or improper bone preparation.
|
| |
|
Search Alerts/Recalls
|