• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. BURR, OVAL, 8 FLUTE 5.0MM X 13CM; BURR,SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. BURR, OVAL, 8 FLUTE 5.0MM X 13CM; BURR,SURGICAL Back to Search Results
Model Number BURR, OVAL, 8 FLUTE 5.0MM X 13CM
Medical Device Problem Code Excessive Heating (4030)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 12/20/2021
Type of Reportable Event Malfunction
Event or Problem Description
On 12/20/2021, it was reported by a sales representative via sems that an ar-8500obe burr got hot instantly after about 15 seconds of being in use.This was discovered during a shoulder arthroscopy/ rotator cuff repair on (b)(6) 2021.Surgeon completed case without further issues using a new burr.
 
Additional Manufacturer Narrative
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.Device discarded.The most likely cause for the reported failure can be attributed to user error of the device due to insufficient fluid flow during use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BURR, OVAL, 8 FLUTE 5.0MM X 13CM
Common Device Name
BURR,SURGICAL
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
emily shafer
8009337001
MDR Report Key19381479
Report Number1220246-2024-03722
Device Sequence Number20436789
Product Code GFF
UDI-Device Identifier00888867043862
UDI-Public00888867043862
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2026
Device Model NumberBURR, OVAL, 8 FLUTE 5.0MM X 13CM
Device Catalogue NumberAR-8500OBE
Device Lot Number14339092
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/20/2021
Initial Report FDA Received Date05/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/15/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-