| Catalog Number |
2162 |
| Medical Device Problem Code |
Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/11/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported a patient underwent an unknown procedure on (b)(6) 2023 and a reservoir was used.The suction reservoir device has jammed.Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Additional information: h6 component code: g07002 - device not returned.Additional information provided: was surgery delayed due to the reported event? unknown.Was procedure successfully completed? unknown.Were fragments generated? unknown.If yes, were they removed easily without additional intervention? unknown, patient status/ outcome / consequences , patient consequence description/was there a clinical outcome experienced by the patient (infection, inflammation, etc.)? unkown.Was other medical intervention (e.G.X-rays, additional procedures, prescriptions, otc, revision) required: unknown.Is the patient part of a clinical study: unknown.Additional information has been requested however not received.Attempts to obtain the device have been made.If further details are received at a later date a supplemental medwatch will be sent.Did this issue contribute to any patient adverse event? were there any patient consequences? is the device available to be returned for evaluation? if so, please provide the status of the return and any available tracking information.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).The device history records were reviewed and the manufacturing criteria was met prior to the release of this lot.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).H3 evaluation: product sample received for analysis.Man during sample evaluation the plate was able to be open and close without any issue.The strap did not interfere on the correct function of the locking mechanism.Therefore, is considered this man factor does not contribute to the reported complaint defect.Materials locking mechanism of reservoir was checked to verify its condition.Sample was evaluated, and during this evaluation it was notice that the latch of plate and its mechanism were in good conditions; the plate was able to be open and close.In addition, plate subassembly needs to be properly closed in order to perform final assembly on finish good.Therefore, is considered this materials factor does not contribute to the reported complaint defect.Based on the present analysis, and condition of sample received it is determined that the reported complaint is not confirmed and it is not related to manufacturing process and other external causes could have affected the product integrity such as handling, non-proper usage or any other possible contributor out of the manufacturing control.The device history records were reviewed and the manufacturing criteria was met prior to the release of this lot.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|