On 03/07/2022, it was reported by a sales representative via sems that qty.2 of an ar-6068-90l 90°, curve, left, quickpass lasso had an issue.The surgeon was performing a posterior labral repair of a right shoulder on (b)(6) 2022.A 90 curve left quickpass lasso was opened to pass sutures in the 6 o'clock position.Out of the package, the scrub tech tried to expel the nitinol guide wire, but it would not come out the tip.The scrub tech then took the nitinol guide wire out of the passer and tried to load the passer again, but again it would not expel out of the tip.After numerous attempts, a second 90 degree curve left quickpass lasso was opened.This lasso had a small amount of nitinol guide wire hanging out the tip, so this was pulled through the tip completely and tried to reload it but the passer once again did not expel the suture.The case was completed successfully by using a conmed spectrum.This was discovered during the unboxing of the complaint device with no case/patient involvement.
|